    Evaluation of Validity of new Biochemistry Analyzer Ark Diagnosis

    The purpose of this study was to evaluate the validation of recently installed biochemistry analyzer ARK Diagnosis FALCON mini to test its analytical performance by two validation parameters Precision and Linearity by mistreatment two analytes glucose and urea. The results obtained were assessed for precision and linearity to determine validity of the instrument. From our findings precision and linearity were within the manufacturer's claims.Our results demonstrate that intraassay precision and linearity of the ARK analyzer meets the requirement of laboratory as per CLSI guidelines. This gives a green signal for installation of the instrument. Maintaining the quality there after is another challenge and an ongoing process maintained by internal and external quality control protocol.